VIAFLO 96/384 – Operating instructions V11 Installation
20 INTEGRA Biosciences AG
3.4.9 Write protect
Select this option to protect programs and menu options from inadvertent modification.
The pipetting programs can still be used.
Keep the password in a safe place. Should you lose your password, contact
INTEGRA Biosciences to retrieve your password.
Write
protect
Description Range
Select an option and press OK to switch protection on or
off:
• Standard Programs
• Custom Programs
• Calibration
• Toolbox
• Password Protection: Protect the access to the write
protect menu by selecting “”.
• Edit Password, if password protect is switched on. To
enter a password use the Touch wheel to highlight a
character and press OK. Press to save the pass-
word. The password must be entered before you can
access the write protect menu.
/
(On/Off)