VIAFLO 96/384 – Operating instructions V11 Maintenance
www.integra-biosciences.com 51
6.5 Calibration
VIAFLO 96/384 can be calibrated by adjusting a correction factor in the software. The
correction factor is determined by a nominal volume and the measured volume. To assess
the accuracy and precision of VIAFLO 96/384 pipetting heads, different methods can be
used:
• Measure one or more channels gravimetrically and then all other channel by photo-
metry. Compare the photometrically determined channels relative to the gravimetri-
cally measured channels. Contact INTEGRA for a detailed protocol.
•Gravimetric only.
• Using commercially available systems (e.g. ARTEL MVS).
INTEGRA Biosciences offers a calibration service for VIAFLO 96/384 pipetting heads.
Please contact your local dealer for more information.
6.6 Equipment disposal
The VIAFLO 96/384 is labeled with the “crossed-out wheeled bin” symbol to
indicate that this device must not be disposed of with unsorted municipal waste.
Instead, it is your responsibility to correctly dispose of your waste equipment by
handing it over to an authorized facility for separate collection and recycling. It
is also your responsibility to decontaminate the device in case of biological,
chemical, and/or radiological contamination so as to protect from health hazards the
persons involved in the disposal and recycling of equipment.
For more information about where you can drop off your waste equipment for recycling,
please contact your local dealer from whom you originally purchased the product or your
local council.
By doing so, you will help conserve natural resources and you will ensure that your waste
equipment is recycled in a manner that protects human health and the environment.
Thank you!

Table of Contents