VIAFLO 96/384 – Operating instructions V11 Operation
28 INTEGRA Biosciences AG
4.4.7 Pipetting between 96 and 384 well plates
For fast and simple reformatting between 96 and 384 well microplates and for loading of
12.5 or 125 µl GripTips with a 96-channel pipetting head, some Plate holders feature a
slide function to shift the microplate in Y-direction beneath the pipetting head. Put this
Plate holder, accommodating the 384 well plate, either on Position A or B (8). Move the
Plate slider (9
) to one of the following 3 positions:
The back position of the Plate slider is used to
accommodate the front positions of 384 well plates.
For pipetting the second columns move the pipetting
head one well to the right, or left respectively.
The front position of the Plate slider accommodates the
back positions of 384 well plates.
For pipetting the second columns move the pipetting
head one well to the right, or left respectively.
Set the Plate slider to the middle to pipet 96 well plates.