EasyManua.ls Logo

Integra VIAFLO 96 - Variable Aspirate Mode

Integra VIAFLO 96
57 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VIAFLO 96/384 – Operating instructions V11 Pipetting modes
38 INTEGRA Biosciences AG
5.2.8 Variable aspirate mode
Application: This mode can be used for a variety of collection applications where the
aspiration volume is well known. This mode is also suited for supernatant collection in
microplates.
Operation:
With the tip(s) in liquid, press and release the RUN key to initiate the first aspiration
volume. Again in liquid, press and release the RUN key to initiate the second aspiration
volume, etc.
Press and hold the RUN key to start Dispense and perform a two-step blowout.
Options Steps Description
Edit Count Sets the total number of aspirating steps.
Asp. Speed Sets speed uniquely for aspirating (1 = slow, 10 = fast).
Disp. Speed Sets speed uniquely for dispensing (1 = slow, 10 = fast).
Aspirate
1...Count
Sets different volumes used for sequentially aspirating (in the
same tip) followed by a single dispense. Number of aspirates
and aspirate volume cannot exceed nominal volume of the
pipetting head.
Speed Sets speed of the current pipetting step.
Custom Converts the predefined program into a custom program.