EasyManua.ls Logo

Integra VIAFLO 96 - Pipetting Specifications

Integra VIAFLO 96
57 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VIAFLO 96/384 – Operating instructions V11 Technical Data
54 INTEGRA Biosciences AG
7.4 Pipetting specifications
Head
Part No.
Volume
Range
(µl)
Min. Scrolling
Volume
Increments (µl)
Test Volume Accuracy Precision
(µl) (±%) (≤%)
6101
6131
0.512.5 0.01
1.25 8.0 3.0
6.25 2.0 1.5
12.5 1.5 1.0
6106
6136
(1 µl–)
1
2–50 0.05
55.02.5
25 2.0 1.5
50 1.5 1.0
6102
6132
(2 µl–)
1
5–125
1. The volumes in brackets refer to extended volumes, see “Pipetting”
under “3.4.4 Preferences” on page 16. For the extended range these
specifications do not apply.
0.1
12.5 3.0 1.7
62.5 2.0 1.5
125 1.5 1.0
6103
(5 µl–)
1
10300
0.5
30 2.5 1.5
150 1.5 1.0
300 1.2 0.8
6104
(25 µl–)
1
501250
1
125 2.5 1.5
625 2.0 1.0
1250 1.8 0.8

Table of Contents