VIAFLO 96/384 – Operating instructions V11 Installation
18 INTEGRA Biosciences AG
3.4.5 Calibration & Services
These options enable you to set calibration features, review service history and move the
pipetting head into the parking position.
After changing desired settings, press to Save.
Calibration
& Services
Description Range
Calibration Allows for re-calibration of VIAFLO 96/384 to restore
accuracy. The calibration factors for Pipet and Repeat
type are displayed.
To edit the calibration volumes, press .
Target Volume: This is the volume you are interested
in using for the calibration.
Actual Volume: This is the measured volume obtained
when dispensing the target volume.
Current Factor: Displays the factor currently in use.
This factor should be the same as specified on the
head or the latest calibration certificate.
Reset: Resets the correction factor back to the origi-
nal factory setting. Press to apply the factory setting.
-
Calibration
Reminder
Sets a calibration reminder based on a number of
pipetting cycles. When the calibration reminder is
displayed, press any key to confirm. However, the
reminder will reappear every time the pipette is turned on
until you change the reminder time or use the reset
option.
Reminder: Press OK to turn the reminder timer On or
Off.
Cycles: Use the Touch wheel to set a reminder inter-
val for calibration (time in thousands of cycles). Press
to set the timer to the defined calibration interval.
Remind in/Total Cycles: Displays the residual amount
of cycles before calibration is required.
Reset: Resets the timer to the defined calibration
interval. Press to enable.
Press to save.
/
(On/Off)
1k - 240k
cycles
Service His-
tory
Displays notes of any service that took place on the
VIAFLO 96/384 listed newest entry first.
-
Park Head Fixes the pipetting and base unit for safe transportation.
Make sure GripTips are ejected, clear the deck and
press the RUN key to start the park routine. The
pipette moves to park position and can then be turned
off.
-