VIAFLO 96/384 – Operating instructions V11 Operation
22 INTEGRA Biosciences AG
4.2.2 Partial tip loading
C
AUTION
If only one tip column should be loaded, reduced loading force is required.
Put a tip box either on the left or right Plate holder (8
). Hold the Control unit (5) and lower
the pipetting head down onto the tip box until the Tip load button (4) flashes.
When prompted to push the Tip load button, first press OK (14
) to switch to “low power”
loading mode. Then push Tip load button while pushing down the Control unit until the
Tip load button lights up.
Move the Pipetting unit (3
) up until the light of the Tip load button turns off.
4.2.3 Ejecting used GripTips
If liquid is in the tips, empty them by pressing the PURGE button (16
). Tips are ejected
by pressing the Tip ejector (18
). Confirm ejection by pressing the Tip ejector a second
time.
When prompted push the Tip load button (a) and at the
same time push down the Control unit (b).
When the tips are loaded, the Tip load button lights up.
Move the Pipetting unit (3
) up until the light of the Tip
load button turns off.
a
b