VIAFLO 96/384 – Operating instructions V11 Pipetting modes
www.integra-biosciences.com 45
As manual custom programs, automatic programs can be created directly on VIAFLO 96/
384. However, we recommend using the VIALINK software to set up automatic custom
programs, see “5.4 VIALINK” on page 46.
Step Description
Blowout / Blowin In automatic mode, the blowout and blowin need to be programmed
after the last dispense.
After a blowout, a blowin has to follow at some point. It does not have
to follow immediately and can have steps in between. E.g. after the
blowout a move step can be programmed to move the tips out of the
liquid, and is then followed by the blowin.
Note: When using “Purge” to empty the tips, a blowout/blowin is
performed automatically and does not need to be programmed.
Delay A delay is a pause between the last and the next step.
If it is set to 0.0 s, pressing the run button is required to trigger the
next step.
Loop A loop repeats the steps between the selected step and the loop
command.
Chose to which step the loop directs and how many times the loop
should be performed.
The number of steps can often be shortened by adding a loop.
Note: Nested loops (loops inside loops) are not allowed.
Tip Change Pauses the program. The user is prompted to eject and load tips
manually. The program continues with the next step after pressing
the RUN key.