Setting the Mode Menu Parameters
1. Simultaneously press the PRINT/ACCUMULATE and ZERO buttons. The display
will show “.”. Press the PRINT/ACCUMULATE button. The flashing display
shows the current setting. Press the ZERO or the LOCAL/TOTAL buttons to toggle
between Auto, on, and off. With ‘Auto’ selected (default), the backlight will
automatically light up when low level light conditions are detected. When the
desired setting is displayed, press the PRINT/ ACCUMULATE button.
2. The display will show “”. Press the PRINT/ACCUMULATE button. The
flashing digit shows the current setting. When the weight displayed is equal to or
greater than the set point, the corresponding icon is displayed on the LCD. During
normal weighing mode, the S1 icon on the display will light when the weight is
greater than or equal to set point 1. Use the LOCAL/TOTAL button the advance the
number and the ZERO button to move the number flashing to the left. When the
display shows the desired number, press the PRINT/ACCUMULATE button.
3. The display will show “”. Press the PRINT/ACCUMULATE button. The
flashing digit shows the current setting. During normal weighing mode, the S2 icon
on the display will light when the weight is greater than or equal to set point 2. Use
the LOCAL/TOTAL button the advance the number and the ZERO button to move
the number flashing to the left. When the display shows the desired number, press
the PRINT/ACCUMULATE button.
4. The display will show “”. Press the PRINT/ACCUMULATE button and the
display will show the current version of firmware loaded in the scale. Press the
PRINT/ACCUMULATE button.
5. The display will show “.”. Press the PRINT/ACCUMULATE button. The flashing
digit shows the current setting. This number is how many readings will be averaged
together before the reading is sent to the display. Higher values will result in a more
stable reading, but will take longer to settle to the final value. Note that the scale
updates at 4Hz, so an Average Rate of ‘8’ equates to 2 seconds of averaging. Enter
a ‘1’ to effectively disable averaging. Use the Lb/Kg button the advance the number
and the ZERO button to move the number flashing to the left. When the display
shows the desired number, press the PRINT / ACCUMULATE button.
6. The display will show “.”. Press the PRINT / ACCUMULATE button. The
flashing digit shows the current Average Threshold setting. This setting enables
dynamic averaging, which can improve the settling time of a large Average Rate. If
the scale senses a large weight change, it will temporarily suspend averaging, jump
to the new weight, and resume averaging. Enter a value of 1-10000 to set the
threshold (in display divisions) at which the dynamic averaging triggers. Enter ‘0’ to
disable dynamic averaging. When disabled, the averaging will never be suspended.
When the display shows the desired number, press the PRINT/ACCUMULATE
button.