LP600 Users Manual
Rev E, February 2017
Page 30 of 43
13.The display will show “”. Apply ¼ of the capacity of the scale load cell 4 using
the same loading fixture. Once the load is applied press the PRINT / ACCUMULATE
button.
14.The display will show “”. Prepare load cell 3 using the loading block and the
rubber pad as per figures 1 and 2 above. With no weight on the scale except the
loading block fixture, press the PRINT / ACCUMULATE button.
15.The display will show “”. Apply ¼ of the capacity of the scale to load cell 3
using the same loading fixture. Once the load is applied press the PRINT /
ACCUMULATE button.
16.The display will show “”. Prepare load cell 4 using the loading block and the
rubber pad as per figures 1 and 2 above. With no weight on the scale except the
loading block fixture, press the PRINT / ACCUMULATE button.
17.The display will show “”. Apply ¼ of the capacity of the scale to load cell 4
using the same loading fixture. Once the load is applied press the PRINT /
ACCUMULATE button.
18.The display will show “” and then return to normal weighing mode. The
cornering has now been internally compensated and saved. Perform corner
calibration verification in normal weighing mode. This concludes the corner
sensitivity adjustment procedure. Return to calibration verification section and
ensure all calibration parameters are met.
Single Corner Adjustment
1. Position the scale, loading block and pad to apply weight on the corner requiring
adjustment.
2. Enter Cal mode.
3. Enter “x” at “” choice to adjust a single corner. The cell number to adjust is
entered in the code as ‘x’. (Example: Enter “” to enter mode to adjust cell #2).
4. The display will appear to return to normal mode, except that now the ‘S1’ and
‘S2’ segments are blinking to indicate you are now in corner adjust mode for your
selected corner.
5. Apply ¼ of the scales capacity to the loading pad over the corner to be adjusted.
6. Press LOCAL/TOTAL to increase or ZERO to decrease the corner’s weight.
(Each press adjusts the display approximately 0.01%)
7. When the desired weight is showing on the display, press PRINT /
ACCUMULATE to save the results and return to normal weighing. Return to
calibration verification section and ensure all calibration parameters are met.