7. Position the corner fixture with rubber pad on the scale as per figure 2. Apply ¼ of
the scale’s capacity to the loading block. Annotate the reading.
8. Repeat for the other five positions.
9. If all readings are within 5 pounds and the previous calibration passed, the
calibration verification is complete. If multiple corner checks fail go to the Multiple
Corner adjustment. If a single corner fails go to Single Corner adjustment.
This concludes the verify calibration procedure.