PASCAL
®
Synthesis Ophthalmic Scanning Laser Systems Page 115 of 141
88292-EN, Rev D
Code
Message
Critical
Clearable
Warning
Description
Action
32
BRF Treatment
Temperature Low
(Moderate)
X
Laser module treatment
temperature is low
Check fan and ventilation.
Allow system to warm up.
33
BRF Treatment
Temperature Low
(Severe)
X
Laser module treatment
temperature is severely low
Check fan and ventilation.
Allow system to warm up.
34
BRF Treatment
Temperature High
(Moderate)
X
Laser module treatment
temperature is high
Check fan and ventilation.
Allow system to cool.
35
BRF Treatment
Temperature High
(Severe)
X
Laser module treatment
temperature is severely high
Check fan and ventilation.
Allow system to cool.
36
LBO Treatment
Temperature Low
(Moderate)
X
Laser module treatment
temperature is low
Check fan and ventilation.
Allow system to warm up.
37
LBO Treatment
Temperature Low
(Severe)
X
Laser module treatment
temperature is severely low
Check fan and ventilation.
Allow system to warm up.
38
LBO Treatment
Temperature High
(Moderate)
X
Laser module treatment
temperature is high
Check fan and ventilation.
Allow system to cool.
39
LBO Treatment
Temperature High
(Severe)
X
Laser module treatment
temperature is severely high
Check fan and ventilation.
Allow system to cool.
40
Red Treatment
Temperature Low
(Moderate)
X
Laser module treatment
temperature is low
Check fan and ventilation.
Allow system to warm up.
41
Red Treatment
Temperature Low
(Severe)
X
Laser module treatment
temperature is severely low
Check fan and ventilation.
Allow system to warm up.
42
Red Treatment
Temperature High
(Moderate)
X
Laser module treatment
temperature is high
Check fan and ventilation.
Allow system to cool.
43
Red Treatment
Temperature High
(Severe)
X
Laser module treatment
temperature is severely high
Check fan and ventilation.
Allow system to cool.
44
Ambient
Temperature Low
(Moderate)
X
The ambient temperature is low
Check fan and ventilation.
Allow system to warm up.
45
Ambient
Temperature Low
(Severe)
X
The ambient temperature is
severely low
Check fan and ventilation.
Allow system to warm up.
46
Ambient
Temperature High
(Moderate)
X
The ambient temperature is high
Allow system to cool.
Reconfigure treatment
parameters if
necessary
47
Ambient
Temperature High
(Severe)
X
The ambient temperature is
severely high
Allow system to cool.
Reconfigure treatment
parameters if
necessary
48
Treatment Base
Temperature Over
Maximum
X
OPSL Base Temperature during
treatment exceeds Laser module
maximum allowable value
Allow system to cool.
Reconfigure treatment
parameters if
necessary
49
Treatment BRF
Temperature Over
Maximum
X
OPSL BRF Temperature during
treatment exceeds
Laser module maximum
allowable value
Allow system to cool.
Reconfigure treatment
parameters if
necessary

Table of Contents