PASCAL
®
Synthesis Ophthalmic Scanning Laser Systems Page 115 of 141
88292-EN, Rev D
BRF Treatment
Temperature Low
(Moderate)
Laser module treatment
temperature is low
Check fan and ventilation.
Allow system to warm up.
BRF Treatment
Temperature Low
(Severe)
Laser module treatment
temperature is severely low
Check fan and ventilation.
Allow system to warm up.
BRF Treatment
Temperature High
(Moderate)
Laser module treatment
temperature is high
Check fan and ventilation.
Allow system to cool.
BRF Treatment
Temperature High
(Severe)
Laser module treatment
temperature is severely high
Check fan and ventilation.
Allow system to cool.
LBO Treatment
Temperature Low
(Moderate)
Laser module treatment
temperature is low
Check fan and ventilation.
Allow system to warm up.
LBO Treatment
Temperature Low
(Severe)
Laser module treatment
temperature is severely low
Check fan and ventilation.
Allow system to warm up.
LBO Treatment
Temperature High
(Moderate)
Laser module treatment
temperature is high
Check fan and ventilation.
Allow system to cool.
LBO Treatment
Temperature High
(Severe)
Laser module treatment
temperature is severely high
Check fan and ventilation.
Allow system to cool.
Red Treatment
Temperature Low
(Moderate)
Laser module treatment
temperature is low
Check fan and ventilation.
Allow system to warm up.
Red Treatment
Temperature Low
(Severe)
Laser module treatment
temperature is severely low
Check fan and ventilation.
Allow system to warm up.
Red Treatment
Temperature High
(Moderate)
Laser module treatment
temperature is high
Check fan and ventilation.
Allow system to cool.
Red Treatment
Temperature High
(Severe)
Laser module treatment
temperature is severely high
Check fan and ventilation.
Allow system to cool.
Ambient
Temperature Low
(Moderate)
The ambient temperature is low
Check fan and ventilation.
Allow system to warm up.
Ambient
Temperature Low
(Severe)
The ambient temperature is
severely low
Check fan and ventilation.
Allow system to warm up.
Ambient
Temperature High
(Moderate)
The ambient temperature is high
Allow system to cool.
Reconfigure treatment
parameters if
necessary
Ambient
Temperature High
(Severe)
The ambient temperature is
severely high
Allow system to cool.
Reconfigure treatment
parameters if
necessary
Treatment Base
Temperature Over
Maximum
OPSL Base Temperature during
treatment exceeds Laser module
maximum allowable value
Allow system to cool.
Reconfigure treatment
parameters if
necessary
Treatment BRF
Temperature Over
Maximum
OPSL BRF Temperature during
treatment exceeds
Laser module maximum
allowable value
Allow system to cool.
Reconfigure treatment
parameters if
necessary