PASCAL
®
Synthesis Ophthalmic Scanning Laser Systems Page 120 of 141
88292-EN, Rev D
to reset.
Reboot.
Call service if error persists.
Invalid Attachment
Calibration
The calibration file is missing, not
readable, or incomplete. Check the
log file for the exact problem.
Reboot.
Call service if error persists.
This warning occurs when a partial
treatment occurs for any reason.
(The most common reason is the
user released the foot switch too
soon, but partial treatments could
also be caused by errors.)
User to select End
Treatment and Return to
the Home Screen.
Re-access treatment screen
and reconfigure treatment
parameters.
If error repeats- End
Treatment and call Service
Error occurs while saving favorite
information to file
Check name
provided. Check for
duplicates.
Power On Self-test failure due to
OPSL laser initialization error
Reboot.
Call service if error persists.
Power On Self-test failure due to
both OPSL laser and one or more
Red lasers initialization error
Reboot.
Call service if error persists.
MM controller
disconnected
Communication with the MM
controller failed.
Reboot.
Call service if error persists.
An invalid 3D controller is
connected.
Power off the system,
disconnect the invalid 3D
controller from the
system and then power
on the system.