EasyManua.ls Logo

KaVo ELECTROmatic TM - Electromagnetic Field Information

KaVo ELECTROmatic TM
90 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for use ELECTROmatic TM and TMM/TMC
2 Safety | 2.8 Electromagnetic fields
2.8 Electromagnetic fields
Electromagnetic fields might interfere with the functions of implanted systems
(such as pacemakers).
Medical electrical devices are subject to special precautions regarding electro-
magnetic compatibility and must be installed and operated in accordance with
the tables of electromagnetic compatibility.
See also:
2 12 Information about electromagnetic compatibility, Page 85
High-frequency communications devices may interfere with medical electrical
devices.
Ask patients if they have a cardiac pacemaker or other system implanted
before you start the treatment.
Comply with the tables of electromagnetic compatibility during installation
and commissioning.
If the device needs to be used in the immediate vicinity of other equipment,
monitor the device or system for malfunctions.
15 / 90

Table of Contents

Related product manuals