Instructions for use ELECTROmatic TM and TMM/TMC
6 Operation | 6.1 Switching the ELECTROmatic on/off
6 Operation
CAUTION
Germ formation.
Infections.
▶ Before treating a patient, let the spray air and spray water exit for at least
20 seconds.
▶ Before the first start-up and after downtimes (weekends, holidays, vaca-
tions, etc.), the air and water lines must be purged and/or rinsed.
▶ The germ reduction of the ELECTROmatic must be carried out via the
dental unit.
See also:
2 8 Reprocessing steps in accordance with DIN EN ISO 17664, Page 74
NOTICE
Incorrectly set input values.
Property damage from incorrect input values.
▶ Check the following input values on the display before use:
▪ Speeds
▪ Transmission ratios
▪ Torque
▪ Direction of motor rotation
▪ SAFEdrive
▪ Settings in ENDO mode
▪ Own settings
NOTICE
Contaminated or moist compressed air at the compressed air connec-
tion
Premature wear
▶ Ensure that the cooling air is dry, clean and uncontaminated in accordance
with EN ISO 7494-2.
Note
Please note the transmission/reduction ratio of the attached handpieces as
these have an impact on the displayed speed and the display torque of the
clamped tool.
6.1 Switching the ELECTROmatic on/off
The unit is ready for operation as soon as it is connected.
▶ To switch the product on, connect the device to the power circuit.
ð The display of the ELECTROmatic shows the input values:
Display ELECTROmatic TM
Display ELECTROmatic TMM/
TMC
Display ELECTROmatic TMC
(COMFORTdrive is connected
and selected)
55 / 90