Instructions for use ELECTROmatic TM and TMM/TMC
6 Operation | 6.2 Starting-up the motor
56 / 90
To switch the device off, unplug the unit from the mains.
Note
The power consumption on idling is so low that the device does not need to
be disconnected.
6.2 Starting-up the motor
Note
The pressure to start the motor is 1 bar/14.5 psi.
The minimum operating pressure at 40,000 rpm is 1.8 bar/26 psi.
Press the foot control until the motor starting pressure (1 bar, 14.5 psi) is
exceeded.
ð The motor/COMFORTdrive starts up.
Push the foot control down as far as it will go.
ð The set maximum speed is reached.
Press the foot control partly down in order to regulate the speed between
minimum and maximum speed.
6.3 Regulating the spray water
CAUTION
Hazard from insufficient amount of spray water.
Damage to the cog due to overheating.
Make sure that the motor speed is appropriate for the preparation on hand.
Use the requisite minimal amount of spray water.
Rotate the regulating sleeve right or left to infinitely adjust the water
portion of the spray.
ð Regulating sleeve rotated all the way to the right = no water supplied
ð Regulating sleeve rotated all the way to the left = maximum amount of
water
1
3
2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KaVo ELECTROmatic TM and is the answer not in the manual?

KaVo ELECTROmatic TM Specifications

General IconGeneral
BrandKaVo
ModelELECTROmatic TM
CategoryDental equipment
LanguageEnglish

Related product manuals