Instructions for use ELECTROmatic TM and TMM/TMC
5 Startup | 5.2 Calibrate the foot control
52 / 90
Affix the cord with cord ties and/or cord tape.
5.2 Calibrate the foot control
Press the foot control down once as far as it will go (maximal pressure) in
order to calibrate the foot control.
ð This starts up the motor and the system calibrates automatically to the ex-
isting system pressure.
Automatic calibration of the foot control
The calibration is performed automatically during the first startup, when the
foot control is pressed for the first time.
Once the calibration of the foot control has been done once, the calibration to
the maximum system pressure takes place automatically in the background
during operation (automatic calibration).
Minor pressure fluctuations are balanced automatically by this measure.
5.3 Measuring the amount of cooling air at the motor
coupling