Instructions for use ELECTROmatic TM and TMM/TMC
8 Reprocessing steps in accordance with DIN EN ISO 17664 | 8.2 Disinfection
8.1.4 Automated external and internal cleaning
Not applicable.
8.2 Disinfection
CAUTION
Germ formation.
Infections.
Before treating a patient, let the spray air and spray water exit for at least
20 seconds.
Before the first start-up and after downtimes (weekends, holidays, vaca-
tions, etc.), the air and water lines must be purged and/or rinsed.
The germ reduction of the ELECTROmatic must be carried out via the
dental unit.
KaVo recommends the following products for germ reduction of the spray water
of the dental unit:
KaVo Oxygenal 6 made by KaVo Dental GmbH www.kavo.com
BluTab made by ConFirm Monitoring Systems Inc www.blutab.com
ICX
®
Water Treatment Tablets made by A-dec
Use in accordance with manufacturer specifications.
Note
The unit must be disinfected manually only.
8.2.1 Manual external disinfection
Use a soft disposable cloth and an approved disinfectant for disinfection by
wiping down all visible surfaces of the unit, foot control surfaces, and con-
necting cables. Make sure that all surfaces are fully wetted.
Let the disinfectant act for the prescribed time.
Dry the surfaces.
Permissible disinfectants (application range in accordance with the available
manufacturer's instructions for use and national guidelines. Please note material
safety data sheets.) KaVo recommends the following products based on the
compatibility of the materials. The microbiological efficacy must be confirmed
by the disinfectant manufacturer.
FD 322 (Dürr)
Microcide AF Liquid (Schülke & Mayr)
CaviCide (Metrex)
Incidin Liquid (Ecolab)
Note
Follow the instructions for use of the disinfectant.
8.2.2 Manual internal disinfection
The germ reduction of the ELECTROmatic must be carried out via the dental
unit.
75 / 90

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KaVo ELECTROmatic TM and is the answer not in the manual?

KaVo ELECTROmatic TM Specifications

General IconGeneral
BrandKaVo
ModelELECTROmatic TM
CategoryDental equipment
LanguageEnglish

Related product manuals