Instructions for use ELECTROmatic TM and TMM/TMC
6 Operation | 6.4 PREP mode
60 / 90
ð SAFEdrive is active = Control element has a blue background.
ð SAFEdrive is deactivated = Control element has a grey background.
ð The settings are always saved instantaneously.
ð In all activated speed memories, the SAFEdrive icon is shown with a blue
background.
Use with the SAFEdrive function
With regard to handpieces from third-party manufacturers, KaVo recommends
inactivating the SAFEdrive function as the SAFEdrive protection function is ef-
fective only with high-speed handpieces from KaVo.
See also:
2 6.4.3.1 Turning SAFEdrive on / off, Page 59
If SAFEdrive is activated and the handpiece is overloaded, the first stage,
"SAFEdrive alert", is activated:
The motor power is reduced (can lead to a drop in motor speed).
The LUX light starts to pulse.
The SAFEdrive icon flashes yellow.
Requirement
The handpiece is not defective
If a handpiece from a third-party manufacturer is used, deactivate SAFEd-
rive or use a high-speed handpiece from KaVo.
If the SAFEdrive icon flashes yellow, relieve the handpiece of its load for at
least 2 seconds, i.e. lift it off the tooth.
ð The motor reaches maximum speed again and the SAFEdrive icon turns yel-
low again.
As soon as the motor reaches maximum speed again and the SAFEdrive
icon turns yellow, continue working as usual making sure to use intermit-
tent mode as the operating mode.
See also:
2 3.6 Technical specifications of the ELECTROmatic, Page 27
If the exposure to the load persists for more than 10 seconds, the stage,
"SAFEdrive error", is reached:
The motor is stopped automatically.
The SAFEdrive icon flashes red.
If a handpiece from a third-party manufacturer is used, deactivate SAFEd-
rive or use a high-speed handpiece from KaVo.
If an automatic motor stop signals that the "SAFEdrive error" status is evid-
ent, remove the handpiece from the mouth of the patient and proceed as
follows:
Carefully check the handpiece head for:
- Temperature
- Damage
- Ability to rotate the dental bur
The motor can be restarted if there is no damage or overheating evident.
To do so, press the foot control again.
When you continue working make sure not to touch the cheek with the
handpiece head and, in particular, with the press-button.