Instructions for use ELECTROmatic TM and TMM/TMC
6 Operation | 6.5 ENDO Mode
▶ Select the torque mode from the selection list.
ð The torque mode is shown on the display.
ð The displayed torque mode is saved.
Torque Control function
Torque Control is active.
▶ Press the foot pedal.
ð The motor starts.
The torque is limited to the set threshold.
Once the limit is reached, the rotation stops, while the torque is being main-
tained.
Once some load is taken off the motor, the rotation continues.
Once 90% of the set torque value are reached, a notifying beep is issued and
the light starts to pulse.
See also:
2 6.5.7 Setting the ENDO torque, Page 70
Note
The ENDO notice signal is emitted once a level of 90% of the set torque is
reached.
Autoreverse function
Autoreverse is active.
▶ Press the foot pedal.
ð The motor starts.
Once 90% of the set torque value are reached, a notifying beep is issued and
the light starts to pulse. Once the set torque is reached, the motor stops and
then automatically starts at the same speed in counterclockwise direction.
▶ To stop this, release the foot pedal.
▶ Press the foot pedal.
ð The motor rotates clockwise again.
▶ To stop this, release the foot pedal.
71 / 90