Instructions for use ELECTROmatic TM and TMM/TMC
6 Operation | 6.5 ENDO Mode
Select the torque mode from the selection list.
ð The torque mode is shown on the display.
ð The displayed torque mode is saved.
Torque Control function
Torque Control is active.
Press the foot pedal.
ð The motor starts.
The torque is limited to the set threshold.
Once the limit is reached, the rotation stops, while the torque is being main-
tained.
Once some load is taken off the motor, the rotation continues.
Once 90% of the set torque value are reached, a notifying beep is issued and
the light starts to pulse.
See also:
2 6.5.7 Setting the ENDO torque, Page 70
Note
The ENDO notice signal is emitted once a level of 90% of the set torque is
reached.
Autoreverse function
Autoreverse is active.
Press the foot pedal.
ð The motor starts.
Once 90% of the set torque value are reached, a notifying beep is issued and
the light starts to pulse. Once the set torque is reached, the motor stops and
then automatically starts at the same speed in counterclockwise direction.
To stop this, release the foot pedal.
Press the foot pedal.
ð The motor rotates clockwise again.
To stop this, release the foot pedal.