:CALibration
:PROTected
:DC
:STEP0
:STEP1
:STEP2
:STEP3 <NRf>
:STEP4 <NRf>
:STEP5 <NRf>
:STEP6 <NRf>
:STEP7 <NRf>
:STEP8 <NRf>
:STEP9 <NRf>
:STEP10 <NRf>
:STEP11 <NRf>
:STEP12 <NRf>
:STEP13 <NRf>
:AC
:STEP1
:STEP2
:STEP3
:STEP4
:STEP5
:STEP6
:STEP7
:STEP8
:STEP9
:STEP10
:STEP11
:STEP12
:STEP13
:STEP14
:STEP15
DC cal steps.
Rear terminal short step.
1
Front terminal short circuit.
Open circuit.
10V DC step.
-10V DC step.
100V DC step.
10 4-wire step.
1k 4-wire step.
10k 4-wire step.
100k 4-wire step.
1M 4-wire step.
10mA DC step.
100mA DC step.
1A DC step.
AC cal steps.
10mV AC at 1kHz step.
100mV AC at 1kHZ step.
100mV AC at 50kHz step.
1V AC at 1 kHz step.
1V AC at 50kHz step.
10V AC at 1kHz step.
10V AC at 50kHz step.
100V AC at 1kHz step.
100V AC at 50kHz step.
700V AC at 1kHz step.
100mA AC at 1kHz step.
1A AC at 1kHz step.
2A AC at 1kHz step.
1V AC at 3Hz step.
1
1V AC at 1kHz step.
1
Notes:
1
DC:STEP0, AC:STEP14, and AC:STEP15 are one-time factory calibration points and are valid only in manufacturing calibration
mode.
2
Upper-case letters indicate short form of each command. For example, instead of sending ":CALibration:PROTected:INITiate,",
you can send":CAL:PROT:INIT.
Table B-1 (cont.)
Remote calibration command summary
Command Description
Calibration Command Reference B-3

Table of Contents

Other manuals for Keithley 2010

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2010 and is the answer not in the manual?

Keithley 2010 Specifications

General IconGeneral
BrandKeithley
Model2010
CategoryMultimeter
LanguageEnglish

Related product manuals