Introduction
This appendix contains detailed information about the various Model 2010 remote calibration
commands. Section 2 of this manual covers detailed calibration procedures. For information
about additional commands to control other instrument functions, refer to the Model 2010 User's
Manual.
Command summary
Table B-1 summarizes Model 2010 calibration commands.
Table B-1
Remote calibration command summary
Command Description
:CALibration
:PROTected
:CODE '<up to 8 char. string>'
:COUNt?
:INITiate
:LOCK
:LOCK?
:SAVE
:DATE <year>, <month>, <day>
:DATE?
:NDUE <year>, <month>, <day>
:NDUE?
Calibration root command.
All commands in this subsystem are protected by the calibration lock
(except queries and :CODE).
Calibration code or password (default: KI002010).
Request the number of times the unit has been calibrated.
Initiate calibration.
Lock out calibration (opposite of enabling cal with :CODE
command).
Request comprehensive cal lock state. (0 = locked; 1 = unlocked)
Save cal constants to EEROM.
Send cal date to 2010.
Request cal date from 2010.
Send next due cal date to 2010.
Request next due cal date from 2010.
B-2 Calibration Command Reference