EasyManua.ls Logo

Keithley 2010

Keithley 2010
134 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
:AC:STEP<n> (CALibration:PROTected:AC:STEP<n>)
Purpose To program individual AC calibration steps.
Format :cal:prot:ac:step<n>
Parameters 1 10mV AC at 1kHz calibration step.
2 100mV AC at 1kHZ calibration step.
3 100mV AC at 50kHz calibration step.
4 1V AC at 1kHz calibration step.
5 1V AC at 50kHz calibration step.
6 10V AC at 1kHz calibration step.
7 10V AC at 50kHz calibration step.
8 100V AC at 1kHz calibration step.
9 100V AC at 50kHz calibration step.
10 700V AC at 1kHz calibration step.
11 100mA AC at 1kHz calibration step.
12 1A AC at 1kHz calibration step.
13 2A AC at 1kHz calibration step.
Description The :AC:STEP command programs the 13 individual AC calibration steps;
<n> represents the calibration step number. The appropriate signal must be
connected to the instrument when programming each step, as summarized in
the parameters listed above.
Example :CAL:PROT:AC:STEP7 Program AC step 7.
Calibration Command Reference B-13

Table of Contents

Other manuals for Keithley 2010

Related product manuals