SCPI Reference Tables 14-11
Table 14-8
STATus command summary
Command Description
Default
Parameter
Ref SCPI
:STATus (Note 1) Sec 15
:MEASurement Measurement event registers:
[:EVENt]? Read the event register. (Note 2)
:ENABle <NRf> Program the enable register. (Note 3)
:ENABle? Read the enable register.
:CONDition? Read the condition register.
:OPERation Operation event registers:
[:EVENt]? Read the event register. (Note 2)
:ENABle <NRf> Program the enable register. (Note 3)
:ENABle? Read the enable register.
:CONDition? Read the condition register.
:QUEStionable Questionable event registers:
[:EVENt]? Read the event register. (Note 2)
:ENABle <NRf> Program the enable register. (Note 3)
:ENABle? Read the enable register.
:CONDition? Read the condition register.
:PRESet Return status registers to default states.
:QUEue Path to access Error Queue:
[:NEXT]? Read the most recent message. (Note 4)
:ENABle <list> Specify error and status messages for Error Queue. (Note 5)
:ENABle? Read the enabled messages.
:DISable <list> Specify messages not to be placed in queue. (Note 5)
:DISable? Read the disabled messages.
:CLEar Clears all messages from Error Queue.
Notes:
1. Commands in this subsystem are not affected by *RST or SYSTem:PRESet. The effects of cycling power, *CLS
and STATus:PRESet are explained by the following notes.
2. Event Registers — Power-up and *CLS clears all bits. STATus:PRESet has no effect.
3. Enable Registers — Power-up and STATus:PRESet clears all bits. *CLS has no effect.
4. Error Queue — Power-up and *CLS clears all bits of the registers.
5. Error Queue Messages — Power-up clears list of messages. *CLS and STATus:PRESet have no effect.

Table of Contents