Remote Operation 11-5
Perform the following steps to select and configure the RS-232 interface:
NOTE To retain a present RS-232 setting, press ENTER with the setting displayed. You can
exit from the menu structure at any time by pressing EXIT.
1. Press SHIFT and then RS232 to access the RS-232 menu. The present state (on or off)
of the RS-232 is displayed.
2. To enable the RS-232 interface:
A. Place the cursor on the on/off selection by pressing the key.
B. Press the or key to toggle the selection to ON.
C. Press ENTER. The present baud rate is displayed.
3. To change baud rate:
A. Place the cursor on the baud rate value.
B. Use the and keys to display the desired baud rate value.
C. Press ENTER. The present flow control setting is displayed.
4. To change flow control:
A. Place the cursor on the present flow control selection.
B. Press the or key to toggle the selection.
C. Press ENTER. The present terminator is displayed.
5. To change the terminator:
A. Place the cursor on the present terminator selection.
B. Press the or key to display the desired terminator.
C. Press ENTER. The instrument returns to the normal display.
NOTE Only the SCPI language can be used with the RS-232 interface. The instrument
defaults to the SCPI language when the RS-232 interface is selected (enabled).