Loading...
Ratio and Delta 5-11
Figure 5-3
Delta measurement connections
Step 3 Configure the trigger model of the SourceMeter.
The menu structure to configure triggers is accessed by pressing CONFIG and then TRIG.
Configure the trigger model as follows:
Arm-Layer:
Arm-In Event IMMEDIATE
Arm-Out TLink Line #3
Arm-Out Events OFF
Arm Count INFINITE
Trig-Layer:
Trigger-In:
Trigger-In Source TRIGGER LINK
Trig-In TLink Line #1
Event Detect Bypass NEVER
Trigger In Events SOURCE=ON DELAY=OFF MEAS=OFF
Trigger-Out:
Trig-Out TLink Line #2
Trigger Out Events SOURCE=ON DELAY=OFF MEAS=OFF
Delay 000.0000 s
Trig-Count 0002 (trigger count must equal the number of sweep points)
DUT
0.1
2182
CH 1
8501 Trigger Link Cable
SourceMeter
HI
LO