Setting limit values
Use the following steps to enter high and low limit values:
1. Press the Limits VALUE key to view the present HI1 limit value:
HI1:+1.000000 ^ (default)
2. To change the HI1 limit, use the cursor keys ( and ) and the manual range keys
(and ) to display the desired value. Move the cursor to the rightmost position (^) and
use the (and ) keys to move the decimal point. Note that with the cursor on the
polarity sign, pressing or toggles the polarity of the value.
3. Press ENTER to view the present LO1 limit value:
LO1:–1.000000 ^ (default)
4. Enter the desired value for this low limit.
5. Press ENTER to view the present HI2 limit value:
HI2:+2.000000 ^ (default)
6. Enter the desired value for this high limit.
7. Press ENTER to view the present LO2 limit value:
LO2:–2.000000 ^ (default)
8. Enter the desired value for this low limit and press ENTER to return to the normal
display.
Enabling limits
Use the following procedure to turn on the limits operation:
1. Press the Limits ON/OFF key to view the present beeper status:
BEEP: NEVER (default)
2. To change the beeper setting, use the and keys to display NEVER, OUTSIDE, or
INSIDE.
3. Press ENTER to return to the normal display. The HI/IN/LO status is displayed along
with the reading.
4. To disable Limits, press the ON/OFF key.
8-4 Limits

Table of Contents

Other manuals for Keithley 2182

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2182 and is the answer not in the manual?

Keithley 2182 Specifications

General IconGeneral
BrandKeithley
Model2182
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals