15-12 Additional SCPI Commands
Figure 15-7
Measurement event enable register
Figure 15-8
Questionable event enable register
RAV
B15 - B10
B9 B8 B7 B6 B5
B4 B3
B2 B1
B0
(2 )
9
(2 )
8
(2 )
7
(2 )
5
(2 )
2
(2 )
1
(2 )
0
32
0/1
Bit Position
Event
Decimal Weighting
Value
Value : 1 = Measurement Event Set
0 = Measurement Event Cleared
LL1
2
0/1
Events : BFL = Buffer Full
BHF = Buffer Half Full
BAV = Buffer Available
RAV = Reading Available
HL2 = High Limit 2
LL2 = Low Limit 2
HL1 = High Limit 1
LL1 = Low Limit 1
ROF = Reading Overflow
BFL
HL1 ROF
512 4 1
0/1 0/1 0/1
BHF BAV
256 128
0/1 0/1
LL2HL2
0/10/1
(2 )
3
8
(2 )
4
16
Cal
B15 - B10
B9 B8 B7 -B5
B4
Bit Position
Event
Decimal Weighting
Value
Value : 1 = Operation Event Set
0=Operation Event Cleared
Events : ACAL = ACAL Summary
Cal = Calibration Summary
Temp = Temperature Summary
Temp
(2 )
4
16
0/1
B3-B0
ACAL
(2 )
9
512
(2 )
8
256
0/1 0/1

Table of Contents

Other manuals for Keithley 2182

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2182 and is the answer not in the manual?

Keithley 2182 Specifications

General IconGeneral
BrandKeithley
Model2182
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals