SCPI Reference Tables 14-13
Table 14-11
Trigger command summary
Command Description
Default
Parameter
Ref SCPI
Sec 7
:INITiate Path to Initiate measurement cycle(s): √
[:IMMediate] Initiate one cycle. √
:CONTinuous <b> Enable or disable continuous initiation. (see Note 1) √
:CONTinuous? Query state of continuous initiation. √
:ABORt Reset trigger system (goes to idle state). √
:TRIGger Path to configure Trigger Layer: √
[:SEQuence[1]] √
:SOURce <name> Select control source; IMMediate, TIMer, MANual,
BUS, or EXTernal.
IMMediate √
:SOURce? Query control source. √
:SIGNal Loop around control source. √
:COUNt <n> Set measure count; 1 to 9999 or INF. (see Note 2) √
:COUNt? Query measure count. √
:DELay <n> Set trigger delay; 0 to 999999.999 (sec). 0 √
:AUTO <b> Enable or disable auto delay. ON √
:AUTO? Query state of auto delay. √
:DELay? Query delay value. √
:TIMer <n> Set timer interval for TIMer control source; 0 to
999999.999 (sec).
0.1 √
:TIMer? Query timer interval. √
:SAMPle Sample Counter:
:COUNt <n> Specify sample count; 1 to 1024. 1
:COUNt? Query sample count.
Notes: 1. Defaults for continuous initiation:
SYSTem:PRESet enables continuous initiation.
*RST disables continuous initiation.
2. Defaults for Trigger Count:
SYSTem:PRESet sets count to INF (infinite).
*RST sets count to 1.