Calibration Reference B-5
:INIT
(:CALibration:PROTected:INITiate)
Purpose To initiate comprehensive and factory calibration procedures.
Format :cal:prot:init
Parameter None
Description The :INIT command enables Model 2700 calibration when performing
these procedures over the bus. This command must be sent to the unit after
sending the :CODE command, but before sending any other calibration
command.
NOTE The :INIT command should be sent only once before performing either DC, AC, or
factory calibration. Do not send :INIT before each calibration step.
Example :CAL:PROT:INIT Initiate calibration.
:LOCK
(:CALibration:PROTected:LOCK)
Purpose To lock out comprehensive or manufacturing calibration.
Format :cal:prot:lock
Parameter None
Description The :LOCK command allows you to lock out both comprehensive and man-
ufacturing calibration after completing those procedures. Thus, :LOCK per-
forms the opposite of enabling calibration with the :CODE command.
NOTE To unlock comprehensive calibration, send the :CODE command. To unlock manu-
facturing calibration, hold in the OPEN key while turning on the power.
Example :CAL:PROT:LOCK Lock out calibration.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD

Related product manuals