3-2 Routine Maintenance
Introduction
The information in this section deals with routine type maintenance and includes procedures
for setting the line voltage, replacing the Model 2700 line and front terminal AMPS fuses, and
replacing the amps fuses for the Models 7700 and 7702 plug-in modules. Replacement of the
Model 2700 non-volatile RAM battery is also covered.
Setting the line voltage and replacing the line fuse
WARNING Disconnect the line cord at the rear panel, and remove all test leads con-
nected to the instrument (front and rear) before replacing the line fuse.
The power line fuse is located in the power module next to the AC power receptacle (see
Figure 3-1). If the line voltage must be changed, or if the line fuse requires replacement, per-
form the following steps:
1. Place the tip of a flat-blade screwdriver into the power module by the fuse holder
assembly (see Figure 3-1). Gently push in and to the left. Release pressure on the
assembly, and its internal spring will push it out of the power module.
2. Remove the fuse, and replace it with the type listed in Table 3-1.
CAUTION For continued protection against fire or instrument damage, replace the
fuse only with the type and rating listed. If the instrument repeatedly
blows fuses, locate and correct the cause of the trouble before replacing the
fuse.
3. If configuring the instrument for a different line voltage, remove the line voltage selec-
tor from the assembly, and rotate it to the proper position. When the selector is installed
into the fuse holder assembly, the correct line voltage appears inverted in the window.
CAUTION Operating the Model 2700 on the wrong line voltage may result in instru-
ment damage.
4. Install the fuse holder assembly into the power module by pushing it in until it locks in
place.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD

Related product manuals