B-18 Calibration Reference
:DATE?
(:CALibration:PROTected:CARD1:DATE?)
(:CALibration:PROTected:CARD2:DATE?)
Purpose To request the Model 7700 calibration date.
Format :cal:prot:card1:date?
:cal:prot:card2:date?
Response <year>, <month>, <day>
Description The :CARD1:DATE? and :CARD2:DATE? queries allow you to read
back the calibration date from a Model 7700 in Slot 1 and Slot 2
respectively.
NOTE The card calibration date is automatically set to the Model 2700 real time clock date
when the card is calibrated.
Example :CAL:PROT:CARD1:DATE? Request card 1 cal date.
:INIT
(:CALibration:PROTected:CARD1:INITiate)
Purpose To initiate Model 7700 calibration procedures.
Format :cal:prot:card1:init
Parameter None
Description The :INIT command enables Model 7700 calibration when performing
these procedures over the bus. This command must be sent to the unit after
sending the :CODE command, but before performing Model 7700
calibration.
Example :CAL:PROT:CARD1:INIT Initiate 7700 calibration.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD