Calibration Reference B-11
:STEP8
(:CALibration:PROTected:DC:STEP8)
Purpose To program the 100k 4-wire comprehensive calibration step.
Format :cal:prot:dc:step8 <Cal_resistance>
Parameter <Cal_resistance> = 90E3 to 110E3 []
Description :STEP8 programs the 100k 4-wire resistance comprehensive calibration
step. The allowable range of the calibration resistance parameter is from
90E3 to 110E3, but 100E3 is recommended for best results.
Example :CAL:PROT:DC:STEP8 100E3 Program 100k step.
:STEP9
(CALibration:PROTected:DC:STEP9)
Purpose To program the 1M comprehensive calibration step.
Format :cal:prot:dc:step9 <Cal_resistance>
Parameter <Cal_resistance> = 900E3 to 1.1E6 []
Description :STEP9 programs the 1M comprehensive calibration step. The allowable
range of the calibration resistance parameter is from 900E3 to 1.1E6. Use
the 1E6 value whenever possible, or the closest possible value.
Example :CAL:PROT:DC:STEP9 Program 1M calibration step.
:STEP10
(CALibration:PROTected:DC:STEP10)
Purpose To program the 10mA comprehensive calibration step.
Format :cal:prot:dc:step10 <Cal_current>
Parameter <Cal_current> = 9E-3 to 11E-3 [A]
Description :STEP10 programs the 10mA comprehensive calibration step. The allow-
able range of the calibration current parameter is from 9E-3 to 11E-3. Use
the 10E-3 value whenever possible for best results.
Example :CAL:PROT:DC:STEP10 10E-3 Program 10mA step.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A