Calibration Reference B-19
:LOCK
(:CALibration:PROTected:CARD1:LOCK)
Purpose To lock out Model 7700 calibration.
Format :cal:prot:card1:lock
Parameter None
Description The :LOCK command allows you to lock out Model 7700 calibration after
completing the procedure. Thus, :LOCK performs the opposite of enabling
calibration with the :CODE command.
Example :CAL:PROT:CARD1:LOCK Lock out card 1 calibration.
:LOCK?
(:CALibration:PROTected:CARD1:LOCK?)
Purpose To read Model 7700 calibration lock status.
Format :cal:prot:card1:lock?
Response 0 Calibration locked.
1 Calibration unlocked.
Description The :LOCK? query requests status from the Model 2700 on Model 7700
calibration locked/unlocked state. Calibration must be enabled sending the
:CODE command before calibration can be performed.
Example :CAL:PROT:CARD1:LOCK? Request card 1 cal lock state.
:RCOunt
(:CALibration:PROTected:CARD1:RCOunt)
Purpose To reset card calibration count to 0.
Format :cal:prot:card1:rco
Parameter None
Description The :RCOunt command resets the card calibration count reported by
:COUNt? to 0.
Example :CAL:PROT:CARD1:RCO Rest card calibration count.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD