EasyManua.ls Logo

Keithley 2700

Keithley 2700
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
B-12 Calibration Reference
:STEP11
(CALibration:PROTected:DC:STEP11)
Purpose To program the 100mA comprehensive calibration step.
Format :cal:prot:dc:step11 <Cal_current>
Parameter <Cal_current> = 90E-3 to 110E-3 [A]
Description :STEP11 programs the 100mA comprehensive calibration step. The allow-
able range of the calibration current parameter is from 90E-3 to 110E-3. Use
the 100E-3 value whenever possible for best results.
Example :CAL:PROT:DC:STEP11 0.1 Program 100mA step.
:STEP12
(CALibration:PROTected:DC:STEP12)
Purpose To program the 1A comprehensive calibration step.
Format :cal:prot:dc:step12 <Cal_current>
Parameter <Cal_current> = 0.9 to 1.1 [A]
Description :STEP12 programs the 1A comprehensive calibration step. The allowable
range of the calibration current parameter is from 0.9 to 1.1. Use a value of
1 whenever possible for best results.
Example :CAL:PROT:DC:STEP12 1 Program 1A step.

Table of Contents

Other manuals for Keithley 2700

Related product manuals