2-14 Calibration
Short and open calibration
1. Connect the Model 8610 low-thermal short to the instrument INPUT and SENSE termi-
nals as shown in Figure 2-1. Make sure the INPUTS button is not pressed in so that the
front inputs are active. Wait at least three minutes before proceeding to allow for ther-
mal equilibrium.
NOTE Be sure to connect the low-thermal short properly to the HI, LO, and SENSE termi-
nals. Keep drafts away from low-thermal connections to avoid thermal drift, which
could affect calibration accuracy.
2. Send the following command:
:CAL:PROT:DC:STEP1
3. After the Model 2700 completes this step, remove the low-thermal short, and then send
this command:
:CAL:PROT:DC:STEP2
NOTE Be sure to minimize movement near front Input terminals. Excessive movements can
cause capacitive coupling errors, which could affect calibration accuracy.
DC volts calibration
After the front panel short and open steps, do the following:
1. Connect the calibrator to the Model 2700 as shown in Figure 2-2. Allow three minutes
for thermal equilibrium.
NOTE Although 4-wire connections are shown, the sense leads are connected and discon-
nected at various points in this procedure by turning calibrator external sense on or
off as appropriate. If your calibrator does not have provisions for turning external
sense on and off, disconnect the sense leads when external sensing is to be turned off,
and connect the sense leads when external sensing is to be turned on.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD

Related product manuals