Calibration 2-23
5. Press SHIFT then TEST, then display TEST:CALIB with the up or down range key.
Press ENTER, select RUN, then enter the appropriate calibration code (default:
002700).
6. Using the up or down range key, select CARD at the CAL:RUN prompt, then press
ENTER.
7. Set the display value to the cold calibration temperature (˚C) you measured in Step 3,
then press ENTER to complete Model 7700 calibration.
Remote Model 7700 calibration
1. Connect the Model 7700 to the 7797 Calibration/Extender Board (see “Extender board
connections” above).
2. With the power off, install the Model 7700/7797 combination in Slot 1, and select the
rear inputs with the INPUTS switch. Allow three minutes for thermal equilibrium.
3. Accurately measure and record the cold temperature of the Model 7700 card surface at
the center of the card.
4. Press in and hold the Model 2700 OPEN key while turning on the power.
5. Enable calibration by sending the :CODE command. For example, the default com-
mand is:
:CAL:PROT:CODE 'KI002700'
6. Initiate calibration by sending the following command:
:CAL:PROT:CARD1:INIT
7. Calibrate the Model 7700 with the following command:
:CAL:PROT:CARD1:STEP0 <temp>
Here <temp> is the cold calibration temperature (˚C) measured in Step 3.
8. Send the following commands to save calibration and lock out calibration:
:CAL:PROT:CARD1:SAVE
:CAL:PROT:CARD1:LOCK

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD