EasyManua.ls Logo

Keithley 6482 - Page 200

Keithley 6482
350 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6482-901-01 Rev. A / August 2012 Return to Section Topics 16-7
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual Section 16: SCPI Command Reference
Table 16-5
SENSe command summary
Command Description
Default
parameter SCPI
[:SENSe[1]]
:CURRent[:DC]
:RANGe
[:UPPer] <n>
[:UPPer]?
:AUTO <b>
:AUTO?
:ULIMit <n>
:ULIMit?
:LLIMit <n>
:LLIMit?
:NPLCycles <n>
:NPLCycles?
:AVERage
:COUNt <n>
:COUNt?
[:STATe] <b>
[:STATe]?
:TCONtrol <name>
:TCONtrol?
:ADVanced
:NTOLerance
<NRf>
:NTOLerance?
[:STATe] <b>
[:STATe]?
:MEDian
:RANK <NRf>
:RANK?
[:STATe] <b>
[:STATe]?
Sense 1 subsystem to control channel 1 measurement:
Path to configure current:
Configure measurement range:
Select range by specifying the expected
current reading; 0 to ±21e-3.
Query range.
Enable or disable autorange.
Query autorange.
Set autoranging upper limit (-0.021 to 0.021).
Query autoranging upper limit.
Set autoranging lower limit (-0.021 to 0.021).
Query autoranging lower limit.
Specify integration rate (in line cycles):
0.01 to 10.
Query integration rate.
Path to configure moving and repeat filters:
Specify filter count; 1 to 100.
Query filter count.
Enable or disable digital filter.
Query state of digital filter.
Select filter control (MOVing or REPeat).
Query filter control mode.
Configure the advanced filter:
Set filter noise window (in %); 0 to 105.
Query filter noise window setting.
Enable or disable advanced filter.
Query state of advanced filter.
Configure and control the median filter:
Specify median filter rank; 1 to 5.
Query median filter rank.
Enable or disable median filter.
Query state of median filter.
2e-4
ON
2e-2
2e-9
1
10
OFF
REPeat
5
OFF
1
OFF
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9

Table of Contents

Related product manuals