MEASure:STRain:QUARter? {<gage_factor> | MIN | MAX | DEF}, [{<range> | MIN | MAX | DEF}, [{<res-
olution> | MIN | MAX | DEF},]]]] (@<ch_list>)
Only supported on DAQM900A, DAQM901A, DAQM902A, and DAQM908A multiplexer modules.
Resets all measurement parameters on the specified channels to their default values. Then, configures the channels
for quarter bridge strain gage measurements and immediately sweeps through the specified channels one time
(independent of the present scan list). The results are sent directly to reading memory and the instrument's output
buffer. This command redefines the scan list.
Parameter Typical return
<gage_factor> = Any numeric value between 0.5 and 5
Default = 2
(none)
<range> = {Auto | 100 mV | 1V | 10 V | 100V | 300V}
Default = AUTO
<resolution> = 1 PLC (0.000003 x Range)
Keysight DAQ970A/DAQ973A Programming Guide 173
2 SCPI Programming