– The CONFigure, MEASure?, [SENSe:]CURRent[:DC]:NPLC, [SENSe:]CURRent[:DC]:RESolution commands auto-
matically disable the aperture mode (these commands select an integration time in NPLCs).
– This parameter is set to its default value after a Factory Reset (*RST command).
[SENSe:]CURRent[:DC]:ZERO:AUTO {OFF | ON} [, (@<ch_list>)]
[SENSe:]CURRent[:DC]:ZERO:AUTO? [(@<ch_list>)]
Enables or disables the autozero mode for DC current measurements.
Parameter Typical return
{OFF | ON}. Default: ON 0 (OFF) or 1 (ON)
Configure DC current measurements and perform autozero:
CONF:CURR:DC 1
CURR:DC:ZERO:AUTO ON
– ON (Default): Internally measures the offset following each measurement. It then subtrascts that measurement
from the preceding reading. This prevents offset readings present on the instrument’s input circuitry from affect-
ing measurement accuracy.
– OFF: The instrument uses the last measured zero measurement and subtracts it from each measurement. It
takes a new zero measurement each time you change the function, range or integration time.
– The autozero mode is set indirectly when you set the resolution and integration time with CONFigure and
MEASure? commands. Autozero is automatically turned OFFwhen you select an integration time less than 1 PLC
with these commands.
– This parameter is set to its default value after a Factory Reset (*RST command).