TRIGger:TIMer {<seconds>|MIN|MAX|DEF}
TRIGger:TIMer? [{MIN|MAX|DEF}]
Sets the trigger-to-trigger interval (in seconds) for measurements on the channels in the present scan list. This com-
mand defines the time from the start of one trigger to the start of the next trigger, up to the specified trigger count
(TRIGger:COUNt command).
Parameter Typical return
<seconds> =
A number from 0 seconds to 360,000 (default: 10) with 1/32,768 Hz resolution
(~30.5 µs)
+1.00000000E+01
Sets the trigger interval to 30 milliseconds.
TRIG:SOUR TIMER
TRIG:TIM 30E-03
– The scan interval applies to the TIMer trigger source as set by the TRIGger:SOURce command.
– If the scan interval is less than the time required to measure all channels in the scan list, the instrument will scan
continuously, as fast as possible (no error is generated).
– The CONFigure and MEASure? commands automatically set the trigger interval to 1 second and the trigger
count to 1 sweep.
– The instrument sets the scan interval to immediate (0 seconds) after a Factory Reset (*RST command).
344
Keysight DAQ970A/DAQ973A Programming Guide
2 SCPI Programming