TRIGger:TIMer {<seconds>|MIN|MAX|DEF}
TRIGger:TIMer? [{MIN|MAX|DEF}]
Sets the trigger-to-trigger interval (in seconds) for measurements on the channels in the present scan list. This com-
mand defines the time from the start of one trigger to the start of the next trigger, up to the specified trigger count
(TRIGger:COUNt command).
Parameter Typical return
<seconds> =
A number from 0 seconds to 360,000 (default: 10) with 1/32,768 Hz resolution
(~30.5 µs)
+1.00000000E+01
Sets the trigger interval to 30 milliseconds.
TRIG:SOUR TIMER
TRIG:TIM 30E-03
The scan interval applies to the TIMer trigger source as set by the TRIGger:SOURce command.
If the scan interval is less than the time required to measure all channels in the scan list, the instrument will scan
continuously, as fast as possible (no error is generated).
The CONFigure and MEASure? commands automatically set the trigger interval to 1 second and the trigger
count to 1 sweep.
The instrument sets the scan interval to immediate (0 seconds) after a Factory Reset (*RST command).