CONFigure:DAC:SENSe (@<ch_list>)
Only supported on DAC output sense channels (channel 06 and 07) in DAQM907A multifunction module.
Configures the DAC output sense channels (channel 06 and 07) in the DAQM907A multifunction module. This com-
mand redefines the scan list.
The CONFigure command does not place the instrument in the "wait-for-trigger" state. Use the INITiate or READ?
command in conjunction with CONFigure to place the instrument in the "wait-for-trigger" state.
Parameter Typical return
(none) (none)
Configure and read a default DACsense measurement on channel 106:
CONF:DAC:SENS (@106)
READ?
Typical Response: +1.32130000E-01
The DACoutput sense channels will make a complementary (voltage or current) measurement based on the out-
put mode of the corresponding DAC output channel.
Channel 06 will make the measurement based on the output mode set in channel 04.
Channel 07 will make the measurement based on the output mode set in channel 05.
This command does makes an actual measurement. The range and resolution are fixed at 1 PLC on the 10
VDCrange.