Description Lower limit Result Upper limit
CV Line Regulation (V
lowline
- V
highline
) -22 mV ______________ 22 mV
CV Ripple and Noise
peak-to-peak
______________ 150 mV
rms
______________ 15 mV
Transient Response ______________ 50 μs
Constant Current Tests
Current Programming
Zero Current Output (I
0
)
-0.0003 A ______________ 0.0003 A
Maximum Current Ouput (I
max
)
0.3995 A ______________ 0.4005 A
Current Readback
Zero Current measured over interface
I
0
- 0.0002 A ______________ I
0
+ 0.0002 A
Maximum Current measured over interface
I
max
- 0.0004 A ______________ I
max
+ 0.0004 A
CC Load Regulation (I
load
- I
short
) -100 μA ______________ 100 μA
CC Line Regulation (I
lowline
- I
highline
) -100 μA ______________ 100 μA
Test Description Instrument settings
Voltage Programming & Readback, Zero Voltage (V
0
) 0 V 0.4 A
Voltage Programming & Readback, Maximum Voltage (V
max
) 100 V 0.4 A
CV Load Regulation, Line Regulation, Ripple and Noise 100 V 0.4 A
Transient Response 100 V 0.2 Ato 0.4 A
Current Programming & Readback, Zero Current (I
0
) 0 A 100 V
Current Programming & Readback, Maximum Current (I
max
) 0.4 A 100 V
CC Load Regulation, Line Regulation, Ripple and Noise 0.4 A 100 V

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E36100 Series and is the answer not in the manual?

Keysight Technologies E36100 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelE36100 Series
CategoryPower Supply
LanguageEnglish

Related product manuals