4- 64 Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
LGI
The LGI command specifies the current monitor channel and its search parameters
for the linear search measurement.
This command ignores the RI command setting.
This command setting is cleared by the LGV command.
Syntax LGI chnum,mode,range,target
Parameters chnum : Search monitor channel number. The value must be slot number where
the module has been installed. Integer expression. 1 to 8 (1 and 5 are
not available for HPSMU). See
Table 4-1 on page 4-9.
mode : Search mode. Integer expression.
0 : If the measured value target, it is the search result data.
1 : If the measured value target, it is the search result data.
range : Measurement ranging type. Integer expression. The measurement range
will be set to the minimum range that covers the target value. For the
limited auto ranging, the instrument never uses the range less than the
specified range. See
Table 4-3 on page 4-11.
target: Search target current (in A). Numeric expression.
0 to ±0.1, 0 to ±0.2 for E5291A, or 0 to ±1 for E5280B/E5290A
Example
Statements
OUTPUT @E5270;"LGI 0,1,14,1E-6"