Keysight E5260/E5270 Programming Guide, Edition 4 4- 65
Command Reference
Command Reference
LGV
The LGV command specifies the voltage monitor channel and its search parameters
for the linear search measurement.
This command ignores the RV command setting.
This command setting is cleared by the LGI command.
Syntax LGV chnum,mode,range,target
Parameters chnum : Search monitor channel number. The value must be slot number where
the module has been installed. Integer expression. 1 to 8 (1 and 5 are
not available for HPSMU). See
Table 4-1 on page 4-9.
mode : Search mode. Integer expression.
0 : If the measured value target, it is the search result data.
1 : If the measured value target, it is the search result data.
range : Measurement ranging type. Integer expression. The measurement range
will be set to the minimum range that covers the target value. For the
limited auto ranging, the instrument never uses the range less than the
specified range. See
Table 4-2 on page 4-10.
target : Search target voltage (in V). Numeric expression.
0 to ±100, or 0 to ±200 for E5280B/E5290A
Example
Statements
OUTPUT @E5270;"LGV 1,2,12,3"

Table of Contents

Other manuals for Keysight Technologies E5260 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E5260 Series and is the answer not in the manual?

Keysight Technologies E5260 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelE5260 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals