Keysight E5260/E5270 Programming Guide, Edition 4 4- 91
Command Reference
Command Reference
PV
The PV command specifies the pulsed voltage source and its parameters. This
command also clears, and is cleared by, the PI command setting.
In the staircase sweep with pulsed bias measurement mode (MM 5 command), the
output forced by the PV command synchronized with the staircase sweep outputs
forced by the WI or WV command.
Measurement channel performs measurement so that the pulse width and pulse
period are kept. The integration time is automatically set by the instrument, and you
cannot change. For the Keysight E5270B, note that the high-resolution ADC cannot
be used for the pulsed measurements. The AAD/AIT/AV/WT command settings are
ignored.
Syntax PV chnum,vrange,base,pulse[,Icomp]
Parameters chnum : Pulsed source channel number. The value must be slot number where
the module has been installed. Integer expression. 1 to 8 (1 and 5 are
not available for HPSMU). See
Table 4-1 on page 4-9.
vrange: Ranging type for the pulsed voltage output. Integer expression. The
output range will be set to the minimum range that covers both base
and pulse values. For the limited auto ranging, the instrument never
uses the range less than the specified range. See
Table 4-4 on
page 4-13.
base,
pulse
: Pulse base voltage or pulse peak voltage (in V). Numeric expression.
See
Table 4-6 on page 4-14.
0 to ±100, or 0 to ±200 for E5280B/E5290A
Icomp: Current compliance value (in A). Numeric expression. See Table 4-6 on
page 4-14. If you do not set Icomp, the previous value is used.
Compliance polarity is automatically set to the same polarity as the
output value, regardless of the specified Icomp. If the output value is 0,
the polarity is set to positive.
Example
Statements
OUTPUT @E5270;"PT 1,0.01"
OUTPUT @E5270;"PV 1,12,0,5,1E-3"
OUTPUT @E5270;"PT 1,0.01"
OUTPUT @E5270;"PV 2,0,0,3"

Table of Contents

Other manuals for Keysight Technologies E5260 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E5260 Series and is the answer not in the manual?

Keysight Technologies E5260 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelE5260 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals