100 Keysight PXIe Chassis Family User Guide
Configuring the PXI Trigger Bus Front Panel Trigger Port Configuration Guidelines
This configuration limitation applies to both directions between Bus 2 and Bus 3.
If you want an input trigger signal to go to both segments, you must check both
boxes. You cannot route the signal between Segment 2 and Segment 3.
This configuration limitation applies to both the GUI and programing interface
using IVI.
The M9018B and the M9019A chassis have three Trigger Bus Segments, which
are numbered 1, 2, and 3. However, the M9010A 10-slot chassis has two Trigger
Bus Segments, which are numbered 1 and 2. With the M9010A, if you want an
input trigger signal to go to both Segments 1 and 2, then you must check both
boxes. You cannot route them between Segments 1 and 2.