100 Keysight NFA Series Noise Figure Analyzers Service Guide
Instrument Messages
Event Messages
–223 Too much data; 200
spurs found.
Additional spurs
ignored.
There are too many spurs for the table (the limit is 200), and any additional
spurs that are found will be ignored
–224 Illegal parameter
value
An exact data value (from a list of the allowed values) was required, but not
found. See the specific feature description for information about the expected
parameter values
–224 Illegal parameter
value; <Value>
invalid. Fractional
values are not
allowed.
The seconds parameter of an LXI time may not contain a fractional portion. For
example 123456789.0 is valid, but 123456789.1 is not valid
–224 Illegal parameter
value; <value> out of
range.
The value does not fall within the valid range
–224 Illegal parameter
value; Exceeding the
max list length
The list parameters have a maximum allowed length. You are trying to set a
length longer than the maximum
–224 Illegal parameter
value; Gated FFT is
not available while
Sweep Type is set to
Swept
The gated FFT function is not available if you have selected the swept type of
sweep. You must select the FFT sweep type
–224 Illegal parameter
value; Gated LO is
not available while
Sweep Type is set to
FFT
The FFT sweep type moves the LO frequency in steps, so the gated LO function
is not available if you have selected FFT sweep
–224 Illegal parameter
value; Gated Video is
not available while
Sweep Type is set to
FFT
The gated video function is not available if you have selected the FFT sweep
type
–224 Illegal parameter
value; Illegal
identifier
<identifier>. This
value may already be
in use.
The value for the LXI LAN identifier parameter must be unique (that is, LAN0
and LAN7 must have different identifier strings)
–224 Illegal parameter
value; Index out of
range
When querying the LXI Event Log or the Servo Log, an index may be used to
look at a specific entry. This error occurs if the index provided does not point to
a valid entry
Err# Message Verbose/Correction Information

Table of Contents

Other manuals for Keysight N8973B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N8973B and is the answer not in the manual?

Keysight N8973B Specifications

General IconGeneral
BrandKeysight
ModelN8973B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals