EasyManuals Logo

Keysight N8973B User Manual

Keysight N8973B
524 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #446 background imageLoading...
Page #446 background image
446 Keysight NFA Series Noise Figure Analyzers Service Guide
Assembly Replacement Procedures
Midplane Assembly
Replacement
1. Refer to Figure 15-43. Install the midplane assembly into the chassis and
attach to the midplane bracket using the six screws (2) removed earlier.
Torque to 9 inch-pounds.
2. Replace the power supply support bracket with the two screws. Torque to
9 inch-pounds.
3. Replace the power supply assembly. Refer to the “Power Supply
Assembly” replacement procedure.
4. Replace the CPU assembly. Refer to the “CPU Assembly” replacement
procedure.
5. Replace the reference assembly. Refer to the “Reference Assembly”
replacement procedure.
6. Replace the top brace and reference bracket. Refer to the “Top Brace”
replacement procedure.
7. Replace the rear panel. Refer to the “Rear Panel” replacement procedure.
8. Replace the instrument outer case. Refer to the “Instrument Outer Case”
replacement procedure.

Table of Contents

Other manuals for Keysight N8973B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N8973B and is the answer not in the manual?

Keysight N8973B Specifications

General IconGeneral
BrandKeysight
ModelN8973B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals