EasyManuals Logo

Keysight N8973B User Manual

Keysight N8973B
524 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #347 background imageLoading...
Page #347 background image
Keysight NFA Series Noise Figure Analyzers Service Guide 347
Service and Diagnostics Menus
Service Key Descriptions
LO Band Lock
Provides the ability to tune the analyzer over as large a range as can be
accommodated by the specified LO Band. As a result, this feature can be used
to check performance in the LO band overlap regions. By definition, multi-band
sweeps are not allowed.
When band lock is in effect the Advisory Event “LO Band Locked: Band <x>”
where “x” is a value between 0 and 6.
When band lock is not in affect the Advisory Event “LO Band Locked: Band
<x>” is cleared.
Dependencies/Couplings:
LO Bandlock will be grayed-out if Bandlock is OFF. If Harmonic Bandlock is any
setting other than OFF, LO Bandlock menu key is enabled. If LO Bandlock is
other than OFF, and the Harmonic Bandlock is transitioned to OFF, the LO
Bandlock is set to OFF.
Off
Turns LO Band Lock to Off. Analyzer will tune from LO band to LO band
normally.
Band 0
Locks the analyzer to LO Band 0.
Band 1
Locks the analyzer to LO Band 1.
Band 2
Locks the analyzer to LO Band 2.
Band 3
Locks the analyzer to LO Band 3.

Table of Contents

Other manuals for Keysight N8973B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N8973B and is the answer not in the manual?

Keysight N8973B Specifications

General IconGeneral
BrandKeysight
ModelN8973B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals